The Effect of Back Reflections on the Acoustic Power Delivered by Physiotherapy Ultrasound Machines

  • Karne A. McBride

      Affiliations

    • Corresponding Author InformationAddress correspondence to: Karne A. McBride, Department of Medical Physics, Western General Hospital, Crewe Road, Edinburgh EH4 2XU, United Kingdom.
  • ,
  • Stephen D. Pye

Department of Medical Physics, Western General Hospital, Crewe Road, Edinburgh, United Kingdom

Received 2 February 2009; received in revised form 30 April 2009; accepted 18 May 2009. published online 17 August 2009.

Abstract 

Physiotherapy ultrasound is used widely for the treatment of soft tissue injuries. The ultrasonic treatment heads are all highly resonant devices and may therefore be sensitive to the levels of acoustic back reflection that they experience. However, the extent to which reflections affect acoustic power during clinical use has not been reported in the literature and is not addressed in current technical standards. This study investigated the effect of physiological levels of acoustic reflection on 29 physiotherapy treatment heads from a total of 21 machines and 6 manufacturers. A range of membranes were constructed and used to mimic the levels of acoustic reflections that occur during treatment. The results obtained showed that almost half of the heads tested (45%) had deviations in acoustic power of more than 15% compared with free-field measurements. Four heads (17%) had deviations in power of more than 25%. We recommend that the susceptibility of physiotherapy ultrasound machines to acoustic reflections be addressed in the relevant technical standards. Also, it is appropriate for tests to be carried out during design and manufacture, and by the purchaser during their acceptance testing. (E-mail: karne.mcbride@luht.scot.nhs.uk)

Key Words: Physiotherapy, Therapeutic ultrasound, Acoustic reflections

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0301-5629(09)00234-8

doi:10.1016/j.ultrasmedbio.2009.05.008

Ultrasound in Medicine and Biology
Volume 35, Issue 10 , Pages 1672-1678, October 2009